Biomechanical insights through finite element analysis of Bennett fracture fixation: a comparative study of four surgical techniques
Journal of Orthopaedic Surgery and Research 2025 Cilt 20 Sayı 90
Scopus Eşleşmesi Bulundu
2
Atıf
20
Cilt
🔓
Açık Erişim
Özet
Background and objectives: Bennett fractures are intra-articular fractures of the first metacarpal, typically resulting from axial loading. The optimal surgical method for fixation remains a topic of ongoing debate. This study aims to evaluate and compare the biomechanical properties of different fixation techniques. Materials and methods: Four fixation models were analyzed: the Lag Screw (LS) model, the Plate-screw (PS) model, and two K-wire models (Kw1 and Kw2). The biomechanical stability of each model was assessed by measuring the total displacement at the fracture line and the maximum von Mises stresses (MvMs) on the fixation materials. Results: Measurements revealed that the LS model exhibited the highest total displacement of 2,184 mm, while the PS model showed the least displacement at 0,069 mm. Among the K-wire models, Kw1 had a total displacement of 0,408 mm, while Kw2 displayed 0,146 mm. Notably, the MvMs values at the fracture line were highest in the LS model (623,95 MPa for the large fragment surface) and lowest in the Kw2 model (16,237 MPa). The PS model demonstrated the lowest stresses on the fixation materials, highlighting its biomechanical advantage. Furthermore, the assessment of joint-level displacement indicated that the Kw1 model had the greatest displacement (0,156 mm), while the PS model maintained remarkable stability, recording a stable 0 mm displacement. Conclusion: The PS model demonstrated superior biomechanical stability, making it the most suitable fixation method among those studied. While the Kw2 model showed promising results with low total displacement values, further studies are necessary to establish its clinical efficacy.
Web of Science Eşleşmesi Bulundu
3
WoS Atıf
20
Cilt
Article
Belge Türü
Kaynak: JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH
Anahtar Kelimeler (WoS)

Havuzumuzdaki Atıflar 0

Bu makaleye, sistemimizdeki Scopus veritabanında bulunan 0 makale atıf yapmıştır. Scopus genel atıf sayısı: 2.

Bu makaleye, kendi Scopus havuzumuzdaki başka bir makaleden atıf kaydı bulunmuyor.
Scimago Dergi Bilgisi Otomatik ISSN Eşleştirmesi 2025 yılı verileri
Journal of Orthopaedic Surgery and Research
Q1
SJR Quartile
1,029
SJR Skoru
84
H-Index
🔓
Açık Erişim
Kategoriler: Orthopedics and Sports Medicine (Q1) · Surgery (Q1)
Alanlar: Medicine
Ülke: United Kingdom · BioMed Central Ltd
Bu bilgiler makale yılına göre Scimago veritabanından ISSN eşleştirmesiyle otomatik getirilmektedir. Dergi sıralama verileri Scimago'nun ilgili yılı baz alınmaktadır.

Anahtar Kelimeler

WoS | Bir kelimeye tıklayıp ilgili kaynaktaki yayınları görün.

Makale Bilgileri

Dergi Journal of Orthopaedic Surgery and Research
ISSN 1749-799X
Yıl 2025 / 1. ay
Cilt / Sayı 20 / 90
Sayfalar 1 – 13
Makale Türü Özgün Makale
Hakemlik Hakemli
Endeks SCI-Expanded
JCR Quartile Q1
Teşvik Puanı 3,60 · YÖKSİS Akademik Teşvik
Yayın Dili İngilizce
Kapsam Uluslararası
Toplam Yazar 5 kişi
Erişim Türü Basılı+Elektronik
Alan Sağlık Bilimleri Temel Alanı Ortopedi ve Travmatoloji

YÖKSİS Yazar Kaydı

Yazar Adı MERCAN NUMAN,ERAVŞAR EBUBEKİR,SAFALI SELİM,UĞUR LEVENT,ÖZDEMİR ALİ
YÖKSİS ID 8718452

Metrikler

Scopus Atıf 2
Havuz Atıfları 0
JCR Quartile Q1
Teşvik Puanı 3,60
Yazar Sayısı 5