Comparison of Immunoassay and Liquid Chromatography-Tandem Mass Spectrometry Methods in the Measurement of Serum Androstenedione Levels
Clinical Laboratory 2018 Cilt 1 Sayı 64
Scopus Eşleşmesi Bulundu
17
Atıf
64
Cilt
69-75
Sayfa
Özet
Background: Recent reports have described inherent problems with androgen immunoassays compared with mass spectrometry analyses. In this study, a new liquid chromatography-tandem mass spectrometry (LC-MS/MS) method was developed according to CLSI rules. The developed method was compared with two immunoassay methods, the enzyme-linked immunosorbent assay (ELISA) and radioimmunoassay (RIA). Methods: After liquid-liquid extraction, a Shimadzu Prominence LC unit coupled to an ABSCIEX API 3200 mass spectrometer with atmospheric pressure chemical ionization was used to quantify serum androstenedione levels. Serum androstenedione results taken from tandem mass spectrometry were compared with the immunoassays. Results: The androstenedione assay was linear up to 50 ng/mL. Lower limit of quantitation and lower limit of detection were 0.195 ng/mL and 0.097 ng/mL, respectively. This method was not affected by matrix effect and other steroid hormone interferences. In this study, the obtained recovery was 69-99%, carryover value was determined as 0.026 ng/mL. According to the results of an interference study, androstenedione bias % did not exceed the limit of allowable bias % and 88.7% recovery was acquired for androstenedione. In the LC-MS/MS and ELISA comparison study, the slope value was found as 18.412, intercept -22.87, and r 2 value as 0.1033. In the LC-MS/MS and RIA comparison study, slope value was found as 1.085, intercept 0.4541, and r 2 value as 0.3712. In the RIA and ELISA comparison study, slope value was found as 9.57, intercept -15.5, and r 2 value as 0.19. Conclusions: The LC-MS/MS provides agreement with the results of radioimmunoassay but not with ELISA. This method offers better selectivity compared to immunoassay systems.
Web of Science Eşleşmesi Bulundu
14
WoS Atıf
64
Cilt
Article
Belge Türü
Kaynak: CLINICAL LABORATORY · s. 69-75
Anahtar Kelimeler (WoS)

Havuzumuzdaki Atıflar 0

Bu makaleye, sistemimizdeki Scopus veritabanında bulunan 0 makale atıf yapmıştır. Scopus genel atıf sayısı: 17.

Bu makaleye, kendi Scopus havuzumuzdaki başka bir makaleden atıf kaydı bulunmuyor.

Anahtar Kelimeler

WoS | Bir kelimeye tıklayıp ilgili kaynaktaki yayınları görün.

Makale Bilgileri

Dergi Clinical Laboratory
ISSN 1433-6510
Yıl 2018 / 1. ay
Cilt / Sayı 1 / 64
Sayfalar 69 – 75
Makale Türü Özgün Makale
Hakemlik Hakemli
Endeks SCI-Expanded
Teşvik Puanı 4,32 · YÖKSİS Akademik Teşvik
Yayın Dili İngilizce
Kapsam Uluslararası
Toplam Yazar 3 kişi
Erişim Türü Elektronik
Alan Sağlık Bilimleri Temel Alanı- Tıbbi Biyokimya

YÖKSİS Yazar Kaydı

Yazar Adı Yücel Kamile, Abuşoğlu Sedat, Ünlü Ali
YÖKSİS ID 3665136

Metrikler

Scopus Atıf 17
Havuz Atıfları 0
Teşvik Puanı 4,32
Yazar Sayısı 3